The AIRSTREAM RELIANT Class II Type A Biological Safety Cabinet is an effective solution in providing operator, product and environmental protection within laboratories and industrial facilities. With the presence of its DC ECM blower, this is the most energy-efficient Class II Biosafety Cabinet in the world with 70% energy savings compared to AC motor. It also features stable and self-compensating airflow, despite building voltage fluctuations & filter loading. Its large performance envelope is an open declaration of possible safe operating airflow values. Certified to EN 12469, Esco AIRSTREAM RELIANT® Class II Biological Safety Cabinet also has antimicrobial coating on all its external and internal painted surfaces for improved safety.
Key Benefits
Inflow | 259 cmh (152 cfm) |
Downflow | 553 cmh (325 cfm) |
Exhaust | 259 cmh (152 cfm) |
Required Exhaust With Optional Thimble Exhaust Collar |
320 cmh (189 cfm) |
Static Pressure For Optional Thimble Exhaust Collar |
29 Pa / 0.11 in H2O |
ULPA Filter Typical Efficiency:
- Downflow >99.999% at 0.1 to 0.3 micron as per IEST-RP-CC001.3 USA
- Exhaust >99.999% at MPPS as per EN 1822 (H-14) EU
Sound Emission (Typical):
- NSF/ANSI 49 56.6 dBA
Fluorescent Lamp Intensity (lux): 1279 lux (119 foot candles)
Cabinet Construction:
- Main Body; 1.2 mm (0.05") 18 gauge electro-galvanized steel with white oven-baked epoxy-polyester Isocide antimicrobial powder-coated finish
- Work Zone; 1.5 mm (0.06") 16 gauge stainless steel, type 304, with BA finish
- Side Walls; UV absorbing tempered glass, 5 mm (0.2"), colorless and transparent
Electrical: 110-120V, AC, 60Hz,1ø
Cabinet Full Load Amps (FLA): 3.5 A
Cabinet Nominal Power: 133 W
Optional Outlets FLA: 5A
Download
Net Weight: 173 Kg (381 lbs)
Shipping Weight: 214 Kg (472 lbs)
Shipping Dimensions, Maximum (W x D x H): 1120 x 820 x 1760 mm (44” x 32” x 69”)
Shipping Volume, Maximum: 1.62 m3 (57 ft3
Esco Airstream Class II Biological Safety Cabinet