Product Description
Yamato SM-530 Large Capacity Steam Sterilizer With Dryer, 50L 220V 20.5-24.5A 50/60Hz
Yamato's large capacity, high performance, fully automatic sterilization from start to finish. The temperature can be adjusted on the easy-to-use 7" touch screen. It's suitable for protein modification at the maximum operating temperature of 135 °C and includes an 11 L heat-resistant container made of stainless steel. The sterilizer is also equipped with safety functions including forced cooling and memory functions and multiple safety-locking mechanisms for the lid. Users can enter sterilization settings (time and temperature) for future use. The unit includes mesh baskets, a drain board, a chemical indicator, a droplet tray, two sterilizing pipes, and two drying pipes.
Operation and functions
-
- Interactive keypad input (touch panel) allows committing sterilization settings (time & temperature) to memory
- 7″ interactive touch screen
- Suitable for protein modification at the maximum operating temperature of 135°C
- Easy settings and operation modes for a multitude of sterilization process
- Increased safety and function list including forced cooling and memory functions
- Equipped with multiple safety locking mechanism for the lid
- Comes with large capacity (11L) heat resistant stainless steel container
- Alarm buzzer sounds when high or low pressure error occurs
-
Specifications
Model | SM-530 | |
System | Automatic high pressure steam sterilization | |
Operating Temperature | Sterilize | 105 to 135°C |
Liquefy | 60 to 110°C | |
Retain | 45 to 60°C | |
Preheat | 45 to 80°C | |
Dry | 135 to 150°C | |
Operating Ambient Temp. | 5 to 35°C | |
Maximum Pressure Capacity | 0.255MPa | |
Heating | Sterilize Pipe | 1000W×2 |
Drying Pipe | 110V/295W×2, 110V/455W×2 | |
Temp. Controller | PID controlled by microprocessor | |
Temp. Setting / Display | Touch panel | |
Timer / Timer Resolution | Range: 0 or 1 min to 99 h 59 min / 1 min | |
Safety Device | Sterilize sensor error, sterilize SSR short circuit, dry sensor error, dry SSR short circuit, sterilize heater disconnection, dry heater disconnection, water level detection (liquid expansion method), independent chamber overheat protection, cover unlock error, chamber over pressure protection, under pressure protection, warning about setting error in cooling water container, memory error, pressure switch (0.25MPa), pressure safety valve (0.255MPa) | |
Internal Dimensions | ⌀370 x D470 mm | |
External Dimensions WxDxH | 520 x 660 x 881 mm | |
Internal Capacity | 50L | |
Power Source (50/60Hz) | Voltage | AC 200-240V (no plug, round terminal) |
Sterilize current | 12.5-10.5A | |
Dry current | 8.0A | |
Weight | ~250lb | |
Included Items | 2 pcs. stainless steel mesh basket (⌀332 x D195.5 mm) | |
OSQ-90 | ||
Drain board x 1, drain bottle x 1, chemical indicator 1 set, filter x 1, droplet tray x 1 |