Product Description
Haier Biomedical DW-86L338J -86°C Upright ULT Freezer 338 L / 11.9 cu ft 115 volt/ 60 Hz
Key Design Features
- Microprocessor-controlled system designed for controllable range of -40°C to -86°C for cabinet space with 1°C increment
- Large LED display for cabinet temperature, set temperature, ambient temperature, and input voltage
- Settable high temperature and low temperature alarms
- Automatic clean-filter alarm and sensor error alert
- Adjustable storage shelf height
- Can be connected with network, and remote alarm systems
- Optional temperature recorder, storage racks and storage boxes
Field Proven Reliability
- Refrigeration system designed and manufactured with best known practice, and components with highest possible quality
- Reliability test performed and passed on all models including aging test on control electronics at 85°C and 85% humidity environment
- Unique insulated inner door design for four separate storage compartments to minimize frost buildup inside the chamber
- Specialized control system design for a well-balanced operation of cascade refrigeration system
Safety
- Equipped with a complete alarm system with optional remote network monitoring system
- UL and CE certified for application safety
- Malfunction alarms including high and low temperature, power outage, sensor errors, high voltage and low voltage, clean-filter, and extremely high ambient, door ajar feature standard on DW-86L486/728/828 models
- Capable of producing two types of alarm outputs: audible buzzer and visible flashing light
- Multiple built-in system protection features including user-settable protection code for controls, user settable delay to start, voltage compensation system, and protection against extreme high voltages
- USB port for temperature data downloading
- Safety padlock compatible on outer door
Installation and Application
- Wide range operating voltage system designed to allow units installed in areas with poor voltage condition
- Suitable for 10°C to 32°C Ambient temperature
- Input voltage and ambient temperature shown simultaneously for ease of monitoring environmental conditions
- Freezer design for global installations
- Robust door latch designed for secure door closing and compact casters for ease of maneuvering
Low Sound Level
- Specialized refrigeration system design using whisper quiet fan and compressors
- Freezer chassis designed to absorb vibration and sound
Energy Saving
- Unique door seal and breaker design for the minimum loss of cold temperature during a door opening
- High performance VIP insulation panels to minimize cabinet heat gain and to improve temperature stability
- Patented cabinet insulation system designed for optimal performance of cold storage temperature and minimal frost buildup
- Condensator fan intelligently start-stop to reduce power consumption
- Unique design of independent insulated inner door systems for independent access of storage space to provide the maximum protection of stored samples and save energy
Warranty Information: 3 years on all parts and 5 years on the compressor.
Optimal Design of Refrigeration SystemCabinet temperature can reach -86℃ with temp uniformity within 6℃
Three independent inner doors and three layer sealing structure, having good heat insulation to reduce energy consumption
Innovative refrigeration cabin design reduces operational noise
Technical Data
Type | -86℃ ULT Freeer |
Cabinet Type | Upright |
Climate Class | N |
Cooling Type | Direct cooling |
Defrost Mode | Manual |
Refrigerant | HC |
Noise Lever (dB) | 50 |
Performance
Cooling performance(℃) | -86 |
---|---|
Temperature Range(℃) | -40~-86 |
Control
Controller | Microprocessor |
Display | LED |
Electrical Data
Power Supply(V/Hz) | 115-60 Hz |
Power(W) | 900 |
Electrical Current(A) | 15 |
Dimensions
Capacity(L/Cu.Ft) | 338/11.9 |
Net/Gross Weight(approx) | 238/278(kg) 524.7/612.9(lbs) |
Interior Dimensions(W*D*H) | 465*630*1165 18.3*24.8*45.9 |
Exterior Dimensions(W*D*H) | 812*893*1846 32.0*35.2*72.7 |
Packing Dimensions(W*D*H) | 875*970*2010 34.4*38.2*79.1 |
Functions
High/Low Temperature | Yes |
---|---|
Remote Alarm | Yes |
Hot Condenser | Yes |
Power Failure | Yes |
High/Low Voltage | N/A |
Sensor Error | Yes |
Low Battery | Yes |
High Ambient Temp | Yes |
Door Ajar | Yes |
Accessories
Caster | Yes |
Foot | Yes |
Test Hole | Yes |
Shelves/Inner Doors | 3/2 |
Temperature Recorder | Optional |
USB Interface | Y |
Rs232/485 Interface | Optional |
CO2 Backup System | Optional |
LN2 Backup System | Optional |
Certificate | CE |
---|