Product Description
Yamato BOA Series Large Capacity Constant Temp Laboratory Oil Bath
BOA201-115V BOA201-220V
Description
Large capacity 37L oil bath with temperature control of up to 200/270°C
Operation and functions
- High temperature distribution accuracy thru jet stirring
- Advanced supportive functions
Standard equipped with external alarm output, temperature output terminal (4~20mA, 1~5V adjustable) RS485 communication function, key lock function, calibration offset function
Safety features
- Triple overheating preventiom function (heater shuts off automatically at set temperature + 6°C, overheating prevention device, independent overheating prevention device)
- Emergency stop switch. Forces cut off of the overcurrent circuit in case of emergency
- Float switch for empty boil prevention and oil overflow prevention
- Operation panel is protected by glass from liquid dripping
- Large operation lamp lighting during operation for attention
- Self diagnostic function, overheat prevention device, over current leakage breaker, key lock function, power failure compensation function
MODEL | BOA201-115V BOA201-220V |
Temp. Control Range* | RT+10~200°C |
Temp. Control Accuracy* | ±0.1°C (at 200°C Silicone Oil) |
Temp. Fluctuation* | 0.2°C (at 200°C Silicone Oil) |
Temp. Distribution Accuracy* | ±0.2°C (at 200°C Silicone Oil) |
Temp. Gradient* | 0.5°C (at 200°C Silicone Oil) |
Temp. Rise Time* | ~120 min. |
Temp. Control System | PID control by microcomputer |
Temp. Setting System | Digital setting with menu keys and the ▼▲ keys |
Temp. Display System | Temp. reading display: Green 4-digit LED digital Temp. setting display: Red 4-digit LED digital |
Sensor | Platinum sensor Pt100Ω (for temperature control) Type-K thermocouple (for overheat prevention) |
Operation Mode | Fixed operation (with operation indicator lamp) |
Stirring Method | Jet agitation |
Heat Insulator | Ceramic fiber |
Heater Type / Capacity | SUS316 Pipe Heater 2kW |
Safety Device | Self-diagnostic functions (automatic overheat prevention, temperature sensor failure, heater disconnection, SSR short circuit, main relay failure, memory error, internal communication error, abnormal temperature reading, abnormal water level), overcurrent leakage breaker, overheating protector, independent overheating preventor, emergency stop button |
Other Functions | Drain valve / operation indicator lamp / external alarm output terminal / temperature output terminal (with 1-5V, 4-20mA changeover switch) / external communication function (RS485) / calibration offset function / set value lock function / power failure recovery mode selection function |
Inner Bath Dimensions | W312 x D440 x H320 mm |
Effective Inner Bath Dimensions | W296 x D340 x H270 mm |
Bath Capacity | ~37L (when the amount of oil is up to 50mm from the upper edge of the bath) |
External Dimensions WxDxH | 531 x 520 x 578 mm (oil bath depth 397 mm) |
Power Source | AC115V 18.5A AC220V 10A |
Weight | ~37kg |
Included Accessories | Shelf 1pc., lid 1pc. |
*1 Performance data above based on 23 ±5 °C room temperature, 65%RH ±20% humidity, and no process load.
Temperature control accuracy, temperature fluctuation, temperature distribution accuracy, and temperature gradient are the values measured using Toshiba Silicone TSF485-50.
Performance varies depending on the environmental temperature, type of medium (water, silicone oil) used, and operating temperature.
*2 Protrusions excluded