Product Description
Labculture® G4 Class II Type B2 (Total Exhaust) Biological Safety Cabinet
The Labculture® G4 Type B2 exhausts 100% of its air through dedicated ductwork with no recirculation, which makes it the right choice when the work involves volatile chemicals or radionuclides alongside biological agents. A fail-safe interlock shuts the main fan down automatically if the building exhaust fails. Both supply and exhaust air pass through ULPA filters, creating an ISO Class 3 work zone, ten times cleaner than the industry-standard ISO Class 5.
This cabinet must be hard-ducted to a dedicated building exhaust system. Use the CBV exhaust volume and static pressure below when sizing your HVAC.
Features
| 100% Total Exhaust No air is recirculated to the room or back into the work zone. All air is exhausted through dedicated ductwork, which makes the cabinet suitable for work with volatile chemicals and radionuclides alongside biological agents. | |
| Fail-Safe Exhaust Interlock If the building exhaust fails, the cabinet’s main fan shuts down automatically to prevent contaminated air from being pushed back into the room. | |
![]() | Centurion Touchscreen Controller Safety information and cabinet alarms shown on a 3D BSC diagram, with clear instructions for each alarm type and a selectable Quickstart mode. Centered and angled down for easy reach and viewing. |
| Dimmable LED Adjustable work zone brightness at an ideal 1200 lux. Saves energy and optimizes work comfort. | |
![]() | Cabinet Construction Stainless steel side walls with side capture zones and negative pressure to optimize containment. |
![]() | Single-Piece Work Tray Large, recessed and spill retaining, with support beams for less vibration and a holder for easy drain pan cleaning. |
![]() | Raised Ergonomic Arm Rest Durable stainless steel arm rest for a comfortable working posture. |
| USB Port and Remote Modbus USB port to access the cabinet data log and connect to your BMS. Remote Modbus lets you reach the cabinet securely from external devices. | |
![]() | Energy-Efficient DC ECM Blower Up to 70% less energy than an AC motor, with airflow that stays stable through building voltage fluctuations and filter loading. |
| ULPA Filter, Supply and Exhaust 10x the filtration efficiency of a HEPA filter on both supply and exhaust. Creates an ISO Class 3 work zone instead of the industry-standard ISO Class 5, with the same 10-year filter life and replacement cost. | |
![]() | Isocide™ Antimicrobial Powder Coating Silver-ion impregnated powder coat eliminates 99.9% of surface bacteria within 24 hours of exposure. |
Specifications
| Nominal Size | 1.8 m (6 ft) |
|---|---|
| External Dimensions, without optional base stand (W × D × H) | 2030 × 867 × 1610 mm · 79.9" × 34.1" × 63.3" |
| Internal Dimensions (W × D × H) | 1873 × 623 × 720 mm · 73.7" × 24.5" × 28.3" |
| Usable Work Area | 0.90 m² · 9.7 ft² |
| Sash Opening | 203 mm · 8" |
| Maximum Sash Opening | 534 mm · 21" |
| Average Inflow Velocity | 0.53 m/s · 105 fpm |
| Average Downflow Velocity | 0.31 m/s · 60 fpm |
| Airflow Volume – Inflow | 724 m³/h · 429 cfm |
| Airflow Volume – Downflow | 1210 m³/h · 700 cfm |
| CBV Exhaust Air Volume | 2142 m³/h · 1277 cfm Use the CBV exhaust volume and static pressure when sizing the building HVAC exhaust and supply. |
| Minimum Exhaust Static Pressure | 325 Pa · 1.3 in. w.g. |
| CBV Exhaust Static Pressure | 498 Pa · 2.0 in. w.g. |
| Supply and Exhaust ULPA Filter Efficiency | >99.999% at 0.1 to 0.3 micron per IEST-RP-CC001.3 (USA) >99.999% at MPPS, H14 per EN 1822 (EU) |
| Work Zone Classification | ISO Class 3 |
| Sound Emission | 60 dBA per NSF/ANSI 49 · 57 dBA per EN 12469 |
| Light Intensity | ≥1200 lux · ≥111 foot-candles, LED lamp |
| Electrical | 115 VAC, 50–60 Hz · Full Load Amps 13.3 A |
| Nominal Power Consumption | 225 W |
| Heat Load | 768 BTU/h |
| Main Body | 1.5 mm / 16 gauge electro-galvanized steel with Isocide™ antimicrobial powder coating |
| Work Zone | 1.5 mm / 16 gauge stainless steel 304, No. 4 finish |
| Sash Window | 6 mm UV-absorbing tempered glass |
| Net Weight | 966 lb |
| Shipping Weight | 1082 lb |
| Shipping Dimensions, maximum (W × D × H) | 2150 × 950 × 1950 mm · 84.7" × 37.4" × 76.8" |
| Shipping Volume | 3.98 m³ · 140.6 cu ft |
| Esco Item Code | 2011363 |
What’s Included
- Esco Labculture G4 Class II Type B2 Biological Safety Cabinet
- Direct-mounted electrical outlet, Plug Code A
- American-style universal service fixture for air, gas, vacuum or nitrogen
- 5-year warranty
Recommended Accessories
| STA-6A0 | Support stand, telescoping 28"–36", caster wheels and leveling feet | $580 |
| UV-30A | UV lamp kit, 253.7 nm | $160 |
| ICO-LB26 | Inlet collar for hard-duct connection, 10-inch diameter | Contact Us |
| IV-1870 | IV bar kit, includes 6 hooks | $220 |
| SF-2U-40 | Universal service fixture kit, 40 mm thread, ball valve | $120 |
| PF-2 | Prefilter, 565 × 500 mm | Contact Us |
Certifications
NSF/ANSI 49 · ISO 14644.1 Class 3 · EN 12469 · EN-1822 (H14) · IEST-RP-CC001.3 · UL 61010-1 · CSA 22.2 No. 61010-1
Do you need total exhaust?
Type B2 is required when volatile chemicals or radionuclides are used in the cabinet. If the work is biological only, the recirculating Type A2 (LA2-6S9-G4) costs less to buy and far less to run, since it does not send conditioned room air out of the building.