Product Description
Spray drying from an organic solvent on an open-cycle machine is not an option: you would be blowing flammable, hot solvent vapour into the room. The GAS-410 is what makes solvent spray drying possible — it closes the loop. The dryer runs on a sealed nitrogen circulation instead of air, and the solvent is condensed out and collected in a 2 L flask instead of being vented.
This is not a standalone unit. It works in conjunction with the ADL-311SA, GB-210A or DL-410. It also removes a purchase: the GAS-410 has a linear compressor built in, so when you run solvent work you no longer need the separate spray-air compressor the dryer would otherwise require.
Key Features
- Sealed nitrogen circulation when connected to an ADL-311SA, GB-210A or DL-410
- Capacitor + freezer recovery system — 400 W air-cooled, fully sealed, R404A
- Solvent collected in a 2 L flask rather than vented
- Integrated linear compressor for spraying — no separate air compressor needed for solvent work
- Roots blower circulation, 0.12 to 0.65 m³/min
- O₂ density monitor with a zirconium limit-current sensor — the number that tells you the loop is genuinely inert
- Filter differential pressure meter so you see the cartridge filter clogging before it becomes a problem
- Cooling trap temperature monitor
- Nitrogen consumption only 15 L/h at 0.1 MPa
Technical Specifications
| Model | GAS-410 (CE) |
| Type | Organic solvent recovery unit — not a standalone unit |
| Compatible with | Yamato ADL-311SA, GB-210A, DL-410 |
| Solvent recovery system | Capacitor + freezer |
| Circulating gas | Nâ‚‚ (sealed circulation when connected to ADL-311SA or GB-210A) |
| Circulating volume flow | 0.12 to 0.65 m³/min |
| Compressor (for spraying) | Linear compressor, integrated |
| Circulation blower | Roots blower |
| Solvent recovery container | 2 L flask |
| Freezer | Air-cooled condensation, fully sealed, 400 W, R404A |
| Filter | Cartridge filter |
| Instrumentation | Cooling trap temperature monitor, filter differential pressure meter, Oâ‚‚ density monitor, blower air volume control |
| Oâ‚‚ sensor | Solid electrolyte (zirconium) limit current type |
| External dimensions (W × D × H) | 700 × 950 × 1500 mm |
| Net weight | approx. 130 kg (287 lb) |
| Power supply (50/60 Hz) | AC 200 to 240 V, 5 A (15 A) |
| Nitrogen required | 15 L/h at 0.1 MPa |
| Country of origin | China |
| Harmonized tariff code | 8419.33.1000 |
| GTIN | 797059929092 |
Safety Features
- O₂ density meter — continuous verification that the circulating loop is inert
- Flammable gas alarm
- Forced Nâ‚‚ introduction when nozzles are removed, so the loop is purged before it is opened
- Electrical leakage breaker
- Anti-seismic clamps supplied
What's Included
- GAS-410 organic solvent recovery unit
- Set of connection parts
- Anti-seismic clamps
- Interface cable
- Sample gas for gas alarm inspection
- 2 L recovery flask
- Instruction manual
Not included: The spray dryer itself, and the nitrogen supply. Nâ‚‚ must be 99.99 % purity (medical grade).
Installation Requirements
- Electrical: AC 200–240 V, 5 A (15 A rated), 50/60 Hz.
- Nitrogen: a regulated supply of 99.99 % (medical grade) Nâ‚‚ at 15 L/h, 0.1 MPa.
- Requires a compatible spray dryer — ADL-311SA, GB-210A or DL-410. This unit does nothing on its own.
- Floor space: 700 × 950 mm, 1500 mm tall, 130 kg — sited next to the dryer.
- Consider the GWS-410 solvent washing unit downstream for contaminants the condenser does not capture.
Compliance
Country of origin: China. Harmonized tariff code 8419.33.1000. One-year manufacturer warranty.
Pro Lab Supply Corp is an authorized Yamato Scientific distributor. Spray drying is an application-specific purchase — tell us your solvent, your feed solids and the particle size you need, and we will specify the unit, the nozzle, and whether you need solvent recovery.